Contact us

Product pricing advice
export@macylab.com
Emergency contact
86-10-15000509891
Leave a message for us
Online Message
Mobile phone browsing

UV-visible spectrophotometer

UV Visible Spectrophotometer

Model:UV-1500CS
Dowload:Download the attachment immediately
Consultation: Consultation
Contact: Contact
Solutions: Solutions

Product description, technical parameters and configuration

ingle-chip microcomputer control, Touch LCD Screen (7 inches)

Large LCD display can display multiple sets of data

Huge memory space to store multiple sets of data and curves

Auto zero,100% adjustment function

Wavelength automatic adjustment

Wrap lights, automatic filter change , large sample pool (5mm ~ 100mm)

Standard samples having a maximum ten-point standard curve measurement functions built

Direct input by K, B factor for quantitative measurement standard curve

Can be directly input and the corresponding standard concentration values establish a standard curve for quantitative measurement

Adjustable electric saving standard curve parameter measuring set, can save a plurality of standard curve

Can be edited at any time of the standard curve , user-friendly

General Parallel printer interface with or slightly beat interfaces can print the title bar, the measurement data , curve parameters , the standard curve and the curve sample points

With USB interface

PC control can be achieved through more accurate and flexible measurement requirements ( standard )


Software features 

Wavelength Scan: the measured object can be 190nm-1100nm full band spectrum scan , can also be segmented scan

Kinetic tests: the time of the measured object scanning , timely detection of changes in concentration caused by changes in absorbance

Quantitative measurement: by establishing a standard curve , the concentration can be measured in the analysis was

Multi- wavelength testing : set a plurality of wavelengths simultaneously , the measured absorbance of a plurality of wavelengths in the test

DNA / protein test: the test formula purity of nucleic acid and protein concentration analysis of DNA / protein


Technical Specifications 

Model

UV1500CS

Wavelength Range

1901100nm

Optical circuit structure

C-T type Parallel double beam

Band Width

2nm

Wavelength Accuracy

±0.5nm

Wavelength Repeatability

≤0.2nm

Photometric Accuracy

±0.3% τ

Photometric Repeatability

0.15% τ

Stray Light

≤0.04% τ (220nm NaI340nm NaNO2)

Stability

0.001A/h@500nm

Photometric Mode

T,A,C,E

Wavelength Setting

Automatic

Photometric Display Range

-0.33A

Display Mode

Touch LCD Screen (7 inches).

Detector

Import Silicon Photodiode

Light Source

Import Deuterium Lamp & Tungsten Halogen Lamp

Power Requirement

AC 220V/50Hz 110V/60Hz

Power

120W

Dimensions(W*D*H)

480×350×220mm

Weight

15Kg