Visible spectrophotometer

Visible spectrophotometer

Model:V-1500
Dowload:Download the attachment immediately
Consultation: Consultation
Contact: Contact
Solutions: Solutions

Product description, technical parameters and configuration

Character


Single-chip microcomputer control, LCD Screen (128*64 Dots)

Large LCD display can display multiple sets of data

Huge memory space to store multiple sets of data and curves

Auto zero,100% adjustment function

Auto wavelength adjustment

Auto filter changing, large sample pool (5mm ~ 100mm)

Standard samples having a maximum ten-point standard curve measurement functions built

Direct input by K, B factor for quantitative measurement standard curve

Can be directly input standard sample and the corresponding standard concentration values and establish a standard curve for quantitative measurement.

Switch off power save standard curve parameter measuring set

General Parallel printer interface with or slightly beat interfaces can print the title bar, the measurement data, curve parameters, the standard curve and the curve sample points

With USB interface

PC control can be achieved through more accurate and flexible measurement requirements (standard) Technical Specifications

 
Technical Specifications
 
Wavelength Range
320~1000nm
Band Width
2nm
Wavelength Accuracy
±0.5nm
Wavelength Repeatability
≤0.2nm
Photometric Accuracy
±0.3% τ
Photometric Repeatability
0.15% τ
Stray Light
≤0.05% τ (340nm NaNO2)
Stability
0.001A/30min@500nm
Photometric Mode
T,A,C,E
Wavelength Setting
Automatic
Photometric Display Range
-0.3~3A
Display mode
LCD Screen (128*64 Dots)
Detector
Import Silicon Photodiode
Light Source
Import Tungsten Lamp
Power Requirement
AC 220V/50Hz 110V/60Hz
Power
120W
Dimensions(W*D*H)
480×350×220mm
Weight
11Kg